British Standard Outlet (Plastic)

MEDICAL GAS OUTLETS
British Standard Gas Outlets

 

Designed to meet the demanding needs of today’s healthcare environments and ensure long-lasting performance and safety in medical gas delivery for hospitals, clinics, and medical universities.

Each gas outlet undergoes comprehensive testing and is clearly marked and colour-coded for easy identification, allowing quick access during critical moments. To deliver gas to a medical device and prevent cross-connections between devices using different gases, gas supply is enabled through non-interchangeable probes with diametric indexing, as recommended in BS 5682:2015.

Compliant with BS 5682:2015, ISO 9170-1:2016

<  Previous - DIN Standard Outlet  | 

|  Next - AGSS Outlet  > (Coming Soon)

  • Secure Connection To guarantee uninterrupted patient care, connections are hermetically sealed to prevent leaks.
  • Easy to Identify ISO-32 colour-coded for quick identification and marked as per ISO 7396-1:2016 for zero mix-ups.
  • Non-interchangeable Connections Smooth, simple action for reliable gas connections with diametrically-indexed probes as specified in BS 5682:2015.
  • Ease of Installation Flared end pipes or hose connections for easy installation and secure connections to the pipeline.
  • Choice of Materials For durability and contamination-free use, specific corrosion-resistant materials suitable for use with medical gases are used in the making of all our gas devices.
  • Pin-indexed Components Each component has gas-specific connections, so there is no chance of cross-connections during routine use or maintenance.

British Standard medical gas outlets are suitable for the supply of

  • Oxygen [O2]
  • Vacuum [Vac]
  • Medical Air [Air-400]
  • Surgical Air [Air-800]
  • Nitrous Oxide [N2O]
  • Carbon Dioxide [CO2]

 

Gas outlets are available in three mounting options for flexible installation in hospital settings.

  • Wall-mounted
  • Bed Head Panel
  • OT and ICU Pendants
 
 
Product Codes
  Wall-mounted Panel-mounted Pendant-mounted
Oxygen AC 1511 AC 1512 AC 1513
Vacuum AC 1521 AC 1522 AC 1523
Air-4 Bar AC 1531 AC 1532 AC 1533
Air-7 Bar AC 1541 AC 1542 AC 1543
Nitrous Oxide AC 1551 AC 1552 AC 1553
Carbon Dioxide AC 1561 AC 1562 AC 1563
Working Pressure
  • 4-5 bar (400-500 kPa)
    [O2, Air-400, Air-800, N2O, CO2]
  • below -0.4 bar (-40 kPa)
    [Vac]
 
Max Inlet Pressure
  • 10 bar (1000 kPa)
    [O2, Air-400, Air-800, N2O, CO2]
  •  5 bar (500 kPa)
    [Vac]
 
Storage Temperature
  • -20ºC to 60ºC
 
Available base block
  • with a flared coupler for brazing
  • with hose connection
British Standard medical gas outlets are designed and manufactured in line with the requirements of
  • ISO 9170-1:2016 / IS 17812-1:2022

  • ISO 7396-1:2016

  • HTM 02-01

 
The outlet-to-device interface has diametric indexing of probes as per
  • BS 5682:2015
 
The product is registered with the Central Drugs Standard Control Organization (CDSCO) under Reference Number:
  • AirCa-Gurga-HR/M/MD/017755

 

See Certificates

Are You Ready to Start With Us?